CLIA-4 CLIA-3 CLIA-2 CLIA-1

As Found

Gravimetric evaluation of pipette performance as received from customer.  Pipette performance prior to disassembling, cleaning, maintenance or calibration.  As Found or As Received test is performed on pipette to verify  accuracy and precision. 

Test Points

Pipette is verified at specific test points as per customer's requirement.  Per ISO 8655 standards Pipettes are verified at 10%(Test Point 1 or TP1), 50%(Test Point 2 or TP2) and 100%(Test Point 3 or TP3) of nominal volumes for any variable volume pipettes.  A fixed volume pipette is always verified at nominal volume.  Verification below 10% is performed only upon special request.

Readings

Desired number of readings at each test point to confirm accuracy and precision.

Cleaning & Maintenance

Inspection, External Cleaning, Complete Disassembly, Internal Cleaning, Parts Performance, Piston Assembly Maintenance, Polishing and Refurbishing, Functional Diagnosis, Re-Assembly, Test for Functionality

As Calibrated

Gravimetric evaluation of pipette after cleaning, maintenance and calibration.   Pipette is calibrated to either manufacturer's or ISO or customer specifications at 10%(Test Point 1 or TP1) and/or checked at (Test Point 2 or TP2) and/or 100%(Test Point 3 or TP3) to verify accuracy and precision .  Service meets or exceeds manufacturer or ISO or customer specifications. 

Test Points

Pipette is verified at specific test points as per customer's requirement.  Per ISO 8655 standards Pipettes are verified at 10%(Test Point 1 or TP1), 50%(Test Point 2 or TP2) and 100%(Test Point 3 or TP3) of nominal volumes for any variable volume pipettes.  A fixed volume pipette is always verified at nominal volume.  Verification below 10% is performed only upon special request.


TP1,TP2,TP3,TP4 TP1,TP2,TP3 TP1,TP3 TP1

Readings

Desired number of readings at each test point to confirm accuracy and precision .

4 4 4 4

Due Date Labels

Next due date labels affixed per customer’s requirement.

Certificate

A NIST traceable certificate of calibration issued in compliance with CLIA/CAP or ISO 17025 standards. 

Warranty

Calibration warranty is valid if pipettes are not abused (physical or chemical trauma), & are used in accordance with the instructions contained in the manufacturer’s operating manual. 

Repairs

Free repair estimates.  Flat fee applies for all major repairs.

Parts

All parts replaced at additional charge.  Seals replaced once a year on certain models and billed to customer.